Transcutaneous Implants in a Porcine Model: The Use of Highly Porous Tantalum
Pages: 32-40 Views: 5623 Downloads: 1544
Authors
[01] Ronald Hugate, The Denver Clinic for Extremities at Risk at Presbyterian/St. Luke’s Medical Center, Denver, Colorado, USA.
[02] Rhonda Clarke, Zimmer Biologics, Austin, Texas, USA.
[03] Tim Hoeman, Product Development and Global Brand Management, Zimmer Trabecular Metal Technology, Parsippany, New Jersey, USA.
[04] Andrea Friedman, Zimmer Trabecular Metal Technology, Parsippany, New Jersey, USA.
Abstract
Porous tantalum has long been used in orthopedics for enhanced bone in-growth. The purpose was to evaluate soft tissue integration into transcutaneous porous tantalum implants. Eighteen porous tantalum and four solid titanium implants were placed within and passed through the subcutaneous tissues of swine. They were harvested after six weeks and examined microscopically to evaluate epidermal contact, soft tissue penetration into and onto the implant, and inflammation. Fifteen of sixteen porous implants demonstrated both soft tissue and vascular tissue penetration. Nine of sixteen porous implants demonstrated epidermal contact with the transcutaneous portion of the implant. A score of 0-4 was used to describe tissue ingrowth (0=none and 4=100%). Soft tissue penetration score averaged 1.25 at the post and 2.63 at the base of the implant. Vascular tissue penetration score averaged 1.0 at the post and 2.63 at the base. Inflammation was evaluated and scored from 0-4 (0=none and 4=marked inflammation). Acute inflammation was present in six of sixteen porous implants with an average score of 1.88 at the transcutaneous post and 0.81 at the subcutaneous implant base. Chronic inflammation was present in every porous implant with a mean score of 1.5 at the post and 1.56 at the base. Three of the four solid titanium implants extruded during the study. The one surviving implant did not demonstrate any epidermal contact on the implant post or tissue ingrowth with chronic and acute inflammation scores of 2 at the base. Porous tantalum transcutaneous implants experienced epidermal, soft tissue, and vascularized tissue adhesion and integration with minimal inflammation. This finding may help prevent deep infection in transcutaneous implants.
Keywords
Transcutaneous Implants, Highly Porous Tantalum, Implant Infection Prevention
References
[01] Belmont PJ Jr, McCriskin BJ, Sieg RN, Burks R, Schoenfeld AJ. Combat wounds in Iraq and Afghanistan from 2005 to 2009. J Trauma Acute Care Surg. 2012; 73(1):3-12.
[02] Bjørg-Tilde S Fevang, Stein A Lie, Leif I Havelin, Lars B Engesæter, Ove Furnes. Improved results of primary total hip replacement. Acta Orthop. 2010; 81(6): 649–659.
[03] Bobyn JD, Stackpool GJ, Hacking SA, Tanzer M, Krygier JJ. Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial. J Bone Joint Surg Br. 1999; 81-B(5): 907-14.
[04] Chimutengwende-Gordon M, Pendegrass C, Blunn G. Enhancing the soft tissue seal around intraosseous transcutaneous amputation prostheses using salinized fibronectin titanium alloy. Biomed Mater 2011; 6: 1-11.
[05] Gawande A. Casualties of war: military care for the wounded from Iraq and Afghanistan. N Engl J Med 2004; 351:2471–5.
[06] Hagberg K, Brånemark R. Consequences of non-vascular trans-femoral amputation: a survey of quality of life, prosthetic use and problems. Prosthet Orthot Int. 2001;25:186–194.
[07] Huang CT, Jackson JR, Moore NB, Fine PR, Kuhlemeier KV, Traugh GH, et al. Amputation: energy cost of ambulation. Arch Phys Med Rehabil. 1979; 60(1):18-24.
[08] Kane RL, Saleh KJ, Wilt TJ, Bershadsky B. The Functional Outcomes of Total Knee Arthroplasty. J Bone Joint Surg Am. 2005; 87(8):1719-24.
[09] Kang NV, Pendegrass C, Marks L, Blunn G. Osseocutaneous integration of an Intraosseous Transcutaneous Amputation Prosthesis Implant used for reconstruction of a transhumeral amputee: Case Report. J Hand Surg 2010; 35A:1130-34.
[10] Koc E, Tunca M, Akar A, Erbil AH, Demiralp B, Arca E. Skin problems in amputees: a descriptive study. Int J Dermatol. 2008; 47(5):463-6.
[11] Levine B, Della Valle CJ, Jacobs JJ. Applications of Porous Tantalum in Total Hip Arthroplasty. J Am Acad Orthop Surg. 2006; 14 (12): 646-55.
[12] Lyon CC, Kulkarni J, Zimerson E, Ross E, Beck MH. Skin disorders in amputees. J Am Acad Dermatol. 2000;42:501–7.
[13] MacKenzie EJ, Jones AS, Bosse MJ, Castillo RC, Pollak AN, Webb LX, et al. Health-care costs associated with amputation or reconstruction of a limb-threatening injury. J Bone Joint Surg Am. 2007;89(8):1685-92.
[14] Meulenbelt HE, Geertzen JH, Jonkman MF, Dijkstra PU. Skin problems of the stump in lower limb amputees: A clinical study. Acta Derm Venereol. 2011; 91(2):173-7.
[15] Pendegrass CJ, Goodship AE, Blunn GW. Development of a soft tissue seal around bone anchored transcutaneous amputation prostheses. Biomaterials. 2006; 27: 4183-91.
[16] Pendegrass CJ, Gordon D, Middleton CA, Sun SN, Blunn GW. Sealing the skin barrier around transcutaneous implants: in vitro study of keratinocyte proliferation and adhesion in response to surface modifications of titanium alloy. J Bone Joint Surg Br. 2008; 90(1):114-21.
[17] Perry EL, Beck PJ, Williams DL, Bloebaum RD. Assessing peri-implant tissue infection prevention in a percutaneous model. J Biomed Mater Res B Appl Biomater. 2010; 92(2): 397-408.
[18] Puckett SD, Lee PL, Ciombor DM, Aaron RK, Webster TJ. Nanotextured titanium surfaces for enhancing skin growth on transcutaneous Osseointegrated devices. Acta Biomaterialia. 2010; 6: 2352-62.
[19] Rosenbaum Chou TG, Petti CA, Szakacs J, Bloebaum RD. Evaluating antimicrobials and implant materials for infection prevention around transcutaneous Osseointegrated implants in rabbit model. J Biomed Mat Res A. 2010; 92(3): 942-52.
[20] Rowe DA, McMinn D, Peacock L, Buis A WP, Sutherland R, Henderson E, et al. Cadence, Energy Expenditure and Gait Symmetry During Music-Prompted and Self-Regulated Walking in Adults with Unilateral Transtibial Amputation. J Phys Act Health. 2013; [epub ahead of print]
[21] Sullivan, Uden M, Robinson KP. Rehabilitation of the Trans-Femoral amputee with an osseointegrated prosthesis: the United Kingdom experience. Orthotics International. 2003; 27: 114-20.
[22] Tillander J, Hagberg K, Hagberg L, Branemark R. Osseointegrated titanium implants for limb prostheses attachments. Clin Orthop Relat Res. 2010; 468: 2781-88.
[23] Williamson DF. Descriptive epidemiology of body weight and weight change in U.S. adults. Ann Intern Med.1993;119:646-49
600 ATLANTIC AVE, BOSTON,
MA 02210, USA
+001-6179630233
AIS is an academia-oriented and non-commercial institute aiming at providing users with a way to quickly and easily get the academic and scientific information.
Copyright © 2014 - American Institute of Science except certain content provided by third parties.